Section 8
Problem/Display screen
Please insert adapter Z.
Color = ***
Error
Selfcheck stopped.
Please check the lamp.
Please close the lid.
Error [xx]
Error
Selfcheck stopped.
Please remove the cuvette.
Please close the lid.
Error
Selfcheck stopped.
Hardware error.
Error [x]
Error
Too much ambient light!
Move device into shade
or close the lid!
All manuals and user guides at all-guides.com
Troubleshooting
Likely Cause
For measurements with 11mm round
cells the adapter Z is required.
Sample coloring outside the measuring
range.
Self-Check Test stops while starting the
instrument
Self-Check Test stops while starting the
instrument
Electronic defect
The instrument sensors detects too
much ambient light.
Action
Insert the adapter Z into cell
compartment #2.
Press OK.
Dilute the sample or select appropriate
color scale.
Check the lamp and replace, if
necessary.
Close the lid.
Press Start Again.
Remove the cuvette/sample cell from
the cell compartment.
Press OK.
Contact the manufacturer or a sales
representative and indicate the error
number
Reduce ambient light. (Avoid direct sun
light.)
Close the lid.
55