1. Press Nominal Values.
2. Press Edit.
An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:
Stray Light
Photometrical accuracy
Wavelength accuracy
3. Press OK when all values are entered and the overview is
displayed.
4. Press Verification.
5. Insert the adapter Z
cell compartment #2.
Advanced Operations
(Figure 4 on page
11) in
39