IEC 60601-1-2 Table 206
Recommended separation distances between portable and mobile RF communication equipment and the
Rocamed MH01
The equipment Rocamed MH01 is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Rocamed MH01 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Rocamed MH01 as recommended below, according to
the maximum power of communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at maximum output power not listed above, the recommended separation distance
d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer
Note 1: at 80 MHz and 800 MHz, the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
NOTE-MH01-FR rev02
Separation distance according to frequency of transmitter
150 KHz to 80 MHz
d=1,17*sqrt (P)
m
0,117
0,370
1,17
3,70
11,7
80 MHz to 800 MHz
d=1,17*sqrt (P)
m
0,117
0,370
1,17
3,70
11,7
800 MHz to 2,5 GHz
d=2,33*sqrt (P)
m
0,233
0,740
2,33
7,40
23,3
70