Calibration selection values for all model:
LBK 3/6a
1Kg/2kg/3kg
2lb/4lb/6lb
• Press [Zero] key, the display will show "Load" then add the selected
calibration mass value to the pan.
• After the calibration mass is added, wait for the stable symbol to come
on and then press [Zero] key, the display will show "PASS", and then
return to normal weighing after 3 seconds.
• If it is outside of the previous calibration value by more than 5% ,it will
show "FAIL L" or "FAIL H" indicating the new calibration weight was too
low(FAIL L) or too high (FAIL H) in respect to the previous calibration
value.

8.0 ERROR MESSAGE

---ADO
© Adam Equipment Company 2014
LBK 6/12a
2Kg/4kg/6kg
4lb/8lb/12lb
---ADL
--ERR-
Unit-E
LBK 12/25a
LBK 30/65a
4Kg/8kg/12kg
10Kg/20kg/30kg
10lb/20lb/25lb
20lb/40lb/60lb
A/D counts are too high
A/D counts are too low
Wrong operation
All the units are disabled
Low battery indication