The following procedures have been validated by
the manufacturer for reprocessing:
Mechanical Disinfection
Pre-cleaning
Cleaning
Disinfection
Drying
Cleaner
Unit
Sterilization
Packaging
Unit
Procedure
Temperature
Holding time
Note: In case of deviating processes, it is the
responsibility of the user to validate the
suitability and effectiveness of the reprocessing!
The handle is designed for up to 1000 reprocessing
cycles and must be replaced every 2 years (for
supplementation: see accessories list / page 2).
Attention: Do not use handles that are damaged
or outdated!
3 min. pre-rinse
5 min at 48 °C, then 2 min cold rinse
and rinse with deionized water.
5 min. at 93 °C
15 min. at 95 °C
Mediclean forte, Dr. Weigert company
WD 290, Belimed company (program
1)
Blue handle in paper/laminate
Sterilizer class B, LISA 522
(S/N 08-0794), W&H company
Fractionated pre-vacuum process
134 °C
18 min.
34