1. Basic performance and safety
If the device* is operated in the electromagnetic environment specified below together with accessories/
spare parts required by the instructions for use, there are no constraints of basic performance or safety.
2. Description of the electromagnetic environment
The device* is intended for use in the electromagnetic environment specified below. The operator or user is
responsible to assure that it is used in such environment only.
3. Emission
Phenomenon/test
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
and flicker
IEC 61000-3-3
4. Immunity
EMC basic standard/
phenomenon
IEC 61000-4-4
Electrical fast transient
(burst)
IEC 61000-4-5
Surge
Compliance
Group 1
Class B
Class A
Passed
Immunity test level
± 2 kV contact
100 kHz
repetition frequency
t = 5/50 ns
± 0,5 kV, ± 1 kV LtL
Grounding conductor
omitted due to protection
class II.
Electromagnetic environment/guidance
The device* is suitable for use in all establish-
ments, including domestic establishments and
those directly connected to the public power
supply network, that supplies buildings used
for domestic purposes.
The device* uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and not likely to cause any interference in
nearby electronic equipment.
Electromagnetic environment/guidance
Mains power quality shall be that of a typical
commercial or hospital environment.
Mains power quality shall be that of a typical
commercial or hospital environment.
- 9 -