EMC basic standard/
phenomenon
IEC 61000-4-6
Conducted disturbances,
induced by RF fields
IEC 61000-4-11
Voltage interruptions
(lowest and highest rated
voltage)
IEC 61000-4-11
Voltage dips (lowest and
highest rated voltage)
WARNING:
Portable RF communication equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the device*, including cables,
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Immunity test level
3 V
0,15 MHz - 80 MHz
6 V in ISM bands
6,765 MHz - 6,795 MHz
13,553 MHz - 13,567 Mhz
26,957 MHz - 27,283 MHz
40,66 MHz - 40,70 MHz
as well as
6 V in amateur radio bands
1,8 MHz - 2,0 MHz
3,5 MHz - 4,0 Mhz
5,3 MHz - 5,4 MHz
7 MHz - 7,3 MHz
10,1 MHz - 10,15 MHz
14 MHz - 14,2 MHz
18,07 MHz - 18,17 MHz
21,0 MHz - 21,4 Mhz
24,89 MHz - 24,99 MHz
28,0 MHz - 29,7 MHz
50,0 MHz - 54,03 MHz
80 % AM at 1 kHz
0 % UT; 250 / 300 cycles
0 % UT; 1 / 2 cycles
at 0, 45, 90, 135, 180,
225, 270 and 315 degrees
0 % UT; 1 cycle
and 70 % UT; 25 / 30
0 % UT; 1 cycle and
70 % UT; 25 / 30 cycles
Single-phase: at 0 degrees
Electromagnetic environment/guidance
When the user of the device* requires
continuous function in the event of disruption
of supply, it is recommended the device
is operated from an uninterruptible power
supply or battery.
Mains power quality shall be that of a typical
commercial or hospital environment.
- 10 -