7.2 Calibration and Test Kits

The following steps must be performed when conducting a calibration or calibration
verification test (bump test) to ensure proper performance of the monitor. Failure to do so
may adversely affect product performance.
When performing a calibration or calibration verification test (bump test) only use
certified calibration gas at the required concentration level.
Do not test with expired calibration gas.
Do not cover or obstruct display or visual alarm cover.
Ensure sensor inlets are unobstructed and are free of debris
Failure to follow instructions outlined in this user manual can result in sickness or death.
When performing a calibration or calibration verification test (bump test) only use certified calibration gas at the
required concentration level. Do not calibrate with expired calibration gas.
If the instrument cannot be calibrated, do not use until the reason can be determined and corrected.
Do not cover or obstruct display or visual alarm cover.
Ensure sensor inlets are unobstructed and is free of debris
A Field Calibration Kit, CM6-FCK, is needed to complete a CO gas test. These are available through local distribution or from
Macurco.
NOTE: CM-xx must be tested at regular intervals in accordance with the requirements of the National Fire Protection
Association (NFPA) 720.
Contents of the FCK
CM6-FCK: (2) Gas Cylinders, (1) 17L 50ppm carbon monoxide (CO) in air and (1) 17L 200 ppm carbon monoxide (CO) in
air
Gas regulator with about two feet of plastic tubing
Humidifier
CM-6-CH calibration hood
FCK Information
Several detectors can be calibrated with one FCK. The only limitation is the amount of gas in the cylinder. The 17-liter cylinder
has approximately 85 minutes of continuous calibration run time. Replacement cylinders are available. The gas cylinder should
be replaced when the pressure gauge on the regulator shows 25-psi or less.
Note: For optimum test results it is suggested that the unit be in clean air, green light on, and be in a low ambient air flow.