Hardness, Total 20 - 500 Mg/L With Tablet - Tintometer Lovibond MD100 Mode D'emploi

Photometer-system
Masquer les pouces Voir aussi pour Lovibond MD100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Methods
GB
tH 2
Zero
Test
tH 2
0.0.0
Zero
!
Test
tH 2
RESULT
Notes:
1.
Strong alkaline or acidic water samples must be adjusted between pH 4 and pH 10
before the tablet is added (use 1 mol/l Hydrochloric acid resp. 1mol/l Sodium hydroxide).
2.
Conversion table:
1 mg/l CaCO
3
1 °dH
1 °fH
1 °eH
MD100_1c 03/2013
Hardness, total with Tablet
20 – 500 mg/l
Fill a clean vial (24 mm Ø) with 1 ml of the water sample and
9 ml deionised water. Close the vial tightly with the cap.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The "Method" symbol flashes for approx. 8 seconds.
The display shows:
Add one HARDCHECK P tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 5 minutes.
(Countdown can be activated, see page 21)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display as CaCO
Tolerances: ± 50 mg/l
mg/l CaCO
3
----
17.8
10.0
14.3
.
3
°dH
°fH
0.056
0.10
----
1.78
0.56
----
0.80
1.43
marks
marks
°eH
0.07
1.25
0.70
----
23