Methods

GB
Mn
Tablet
Powder
0.0.0
Tablet
Powder
Zero
!
Test
Mn
Tablet
Powder
RESULT
Tablet
Powder
Reagent
Set
MANGANESE LR No. 1 / No. 2
MANGANESE LR No. 1
MANGANESE LR No. 2
28

Manganese with Tablet

0.2 – 4.0 mg/l Mn
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one MANGANESE LR 1 tablet straight from the foil to the
10 ml water sample and crush the tablet using a clean stirring rod.
Dissolve the tablet.
Add one MANGANESE LR 2 tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablets are dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 5 minutes.
(Countdown can be activated, see page 27)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Manganese.
Tolerance: ± 0.3 mg/l Mn
Conversion:
mg/l MnO
= mg/l Mn · 2.17
4
mg/l KMnO
= mg/l Mn · 2.88
4
Form of reagent/Quantity
Tablet / per 100
inclusive stirring rod
Tablet / 100
Tablet / 100
marks
Order-No.
4517621BT
4516080BT
4516090BT
AL100_7a 05/2019