Oxygen, Dissolved With Vacu-Vials - Tintometer Lovibond MD100 Mode D'emploi

Photometer-system
Masquer les pouces Voir aussi pour Lovibond MD100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 72
Methods
GB
O2
0.0.0
Zero
Test
O2
RESULT
56

Oxygen, dissolved with Vacu-vials

10 – 800 μg/l O
2
Insert the adapter for 13 mm Ø round vials.
Place the blank in the sample chamber and perform zero calibration
(see "Operation"). The blank is part of the test kit.
Water should flow through the special sample container for several
minutes to remove any air bubbles sticking at the surface.
The water must flow from the bottom to the top.
When the sample container is bubble-free press one Vacu-vial
the lower edge of the sample container. The Vacu-vial
neck and the vial fills automatically.
A small volume of inert gas remains in the Vacu-vial
Remove the Vacu-vial
immediately.
As the contents of the vial has a higher density than water, it is impor-
tant to remove the vial from the sample container within 5 seconds to
prevent any loss of reagent.
Invert the vial several times. Dry the outside of the vial.
Place the Vacu-vial
in the sample chamber.
®
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in µg/l Oxygen.
point downwards from the sample container
®
K-7553
®
into
®
breaks at the
®
.
®
MD100_5a 08/2016

Publicité

Table des Matières
loading

Table des Matières