rainbow® Extension Cable
RC-EXT
Reusable
Prior to using this Cable the user should read and understand the Operator's Manual for the Device or
Monitor, the Directions for Use for the Sensor and Cable and this Directions for Use.
DESCRIPTION
The rainbow® Extension Cable, RC-EXT, is designed for use with Masimo M-LNC or RC patient cables on
Masimo or Masimo compatible pulse oximetry systems. When used with M-LNC patient cables, the RC-EXT is
able to provide Masimo SET® (SpO
to provide Masimo rainbow SET™ (SpO
WARNINGS
• Always refer to the oximeter module operator's manual for complete instructions or additional
instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the
cable and discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor,
cable and sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent
readings, inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or
disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to
sterilize the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes
may damage the electrical components, potentially leading to patient harm.
INSTRUCTIONS
NOTE: The maximum length for the sensor, cable and extension cable combination is 15 feet.
Connecting
to the Patient Cable
Refer to Fig. 1A. RC-EXT with protective cover: Properly orient the RC-EXT connector  (1), insert the
patient cable connector (2) to the RC-EXT connector and completely close the protective cover of the
RC-EXT (3).
Refer to Fig. 1B. RC-EXT without protective cover: Properly orient the RC-EXT connector and insert the
patient cable connector to the RC-EXT connector.
DIRECTIONS FOR USE
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
and PR) parameters. When used with RC patient cables, the RC-EXT is able
2
, PR, SpCO®, SpMet®, SpHb®) parameters.
2
3
en
Non-Sterile
36179/7692F-0418