Electrical installation
The product is to be integrated in the framework of the measures for lightning
and overvoltage protection at the location of the operator.
Connection cable between the display unit type LWG 2000 and the probe
Cable cross section
Design
Maximum length
Connection
Electrical installation
Connection line between the indicator type LWG 2005 and the probe
Line cross-section
Version
Maximum length
Connection
During connection, additionally secure the indicator type LWG 2000 and 2005 with a
back-up fuse
Connection in indicator
type LWG 2000
Connection in indicator
type LWG 2005
Following the connection of the terminals, replace the housing cover!
8 / 10
2 x 0.5mm²
Wet room NYM or YR, in soil NYY or similar
100m implement cross section of cable as 2 x 1.5mm²
Display unit: Terminals 4 and 5 "Fühler [Probe]"
Probe cable: on the display unit or extension via cable extension
fitting accessory (part no. 15 379 00).
2 x 0.5 mm²
Wet room NYM or YR, in soil NYY or similar
Configure 100 m line cross-section in 2 x 1.5 mm²
Indicator: "Tank 1 to tank 5" terminals
Pull the probe cable through the cable sleeve and connect to the
appropriate terminal. The polarity is unimportant.
Probe cable on the indicator or extension via cable connection
fitting accessory (order no. 15 379 00).
Terminals 1 + 3
Terminals 2 + 3
Alarm terminal P and Ö
Alarm terminal P and Ö
closed in case of "alarm"
open in case of "alarm"
open in case of "alarm"
closed in case of "alarm"
part no. 15 073 96
Probe