English
Calibrating the K-Type sensors (using room temperature)
1. Switch off the DIGIMON
2. Plug in the thermocouples (K-type).
Room and measurement temperatures must remain constant at 25 +/-3°C and
be stable for at least 20 minutes. The thermocouple cables must be straight
(unwound).
During calibration, you should neither touch the cables nor the plugs.
3. While holding down the
pressing the
display shows, "Password" 000
4. Press the
5. Press the
Display shows "T-call"
6. Press the
temperature sensor/clamp calibration mode.
7. During calibration, ensure that both thermocouples are well connected to the
DIGIMON.
The display shows
LP
00xxx
27.0°C
The current temperature displayed on the DIGIMON should not deviate by
more than +/-0.5°C from the room temperature.
8. Press the
The display shows:
LP
End
00xxx
27.0°C
18
LIGHT
ON/OFF
button once. Continue holding down the Light button until the
+
-
R
R
/
buttons to scroll to code 018.
SET
button once.
-
R
button once and then the
SET
button to confirm.
Instruction Manual DIGIMON-SE / DIGIMON4
button continuously, switch on the device by
SET
(low pressure side)
Temperature code
Current temperature
button once to enter