Auto Zero Mode
D-TEK CO
by default will automatically zero to the background, allowing you to
2
search for a higher concentration and help pinpoint the leak source. For example,
if a leak is detected and D-TEK-CO
alarm will stop. This indicates the unit has zeroed to the new background. It will
now take a larger concentration of CO
clean air (no CO
) for several seconds will reset the zero point to normal.
2
Manual Zero Mode
D-TEK CO
can be changed to manual mode, which disables the auto zero
2
feature and allows the user to manually zero to the current background.
NOTE: For best performance, allow D-TEK CO
before enabling manual zero mode.
1. To enable manual zero mode, press and hold the HI/LO button for 5 seconds.
The HI and LO indicators will flash quickly for several seconds followed by a
steady flashing of the HI indicator when manual zero mode is active.
2. Press the HI/LO button to zero to the current background. The HI and LO
indicators will flash quickly while D-TEK CO
NOTE: While in manual zero mode, it is best to periodically zero to the current
background even if a leak is not detected.
3. To return to auto mode, press and hold the HI/LO button until the HI indicator
becomes steady.
NOTE: When D-TEK CO
Changing the Filter Cartridge
D-TEK CO
utilizes a specially designed filter cartridge that fits into a protective
2
cap. The filter cartridge should be changed when it appears dirty or when
substances trapped in the filter cartridge seem to be affecting the sensitivity of
D-TEK CO
.
2
NOTE: Water or oil will not penetrate the filter material, but will prevent airflow
through the filter and affect sensitivity.
Always operate D-TEK CO
Operating without a filter can cause serious damage to the
instrument.
1. Ensure D-TEK CO
2. With the probe tip pointing downward, unscrew and remove the probe cap.
NOTE: Do not allow moisture or dust to get into the probe.
is held in that area for several seconds, the
2
to cause an alarm. Holding D-TEK CO
2
is turned ON, it will always default to auto zero mode.
2
is OFF.
2
to warm up for 5 to 10 minutes
2
zeros.
2
with a filter cartridge installed.
2
6
in
2