5. Reset the fuse seat
Replace the fuse seat in the power socket. Push the fuse case by using the screwdriver, and
turn it clockwise, the fuse seat will be locked when released.
6. Switch on the power
Plug the socket and switch on the power.
Replace lamps
Hot!
Wait 20 minutes before open the lamp chamber after power off to avoid
scald!
1. Tools preparation
Prepare a 6×150mm Flat Blade screwdriver and a pair of glove.
2. Power Off
Switch off the power supply and unplug the socket.
3. Open the cover
Loosen the indicated two screws and remove the lamp cover.
4. Replace the D2 lamp
If your spectrophotometer is visible model, please skip to step 5.
Unplug the connector (No. 2) .Unscrew the 2 screws on the D2 Flange (No.1) and remove
the D2 lamp. Draw on the cotton glove and replace a new lamp. Fix the 2 screws and plug
the connector again.