Verifying Stray Light; Verifying Noise; Verifying Dark Noise; Verifying Stability - Onda V-11 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Verifying Stray Light

EN
Select Tab Stray light in the Performance verification interface.
Reference
Material:
Measurement:

Verifying Noise

Select Tab Noise in the Performance verification interface.
Reference
Material:
Measurement:

Verifying Dark Noise

Select Tab Dark Noise in the Performance verification interface.
Reference
Material:
Measurement:

Verifying Stability

Select Tab Noise in the Performance verification interface.
Reference
Material:
Measurement:

Verifying Bandwidth

Select Tab Dark Noise in the Performance verification interface.
Reference
Material:
Measurement:
10g/l NaI solution or equivalent filter (220nm, UV/VISIBLE models only),
50g/L NaNo2 solution or equivalent filter (340 or 360nm)
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength of
measurement;
2.
Put the Reference in the measurement channel, press the button Zero;
3.
Put the Standard Sample in the measurement channel, press the
button Measure, the result is the stray light of this wavelength.
None
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength of
measurement, press the button Zero;
2.
Press the button Measure, the result is the noise of this wavelength.
Block
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength of
measurement, press the button Zero;
2.
Put the Block in the measurement channel, press the button Measure,
the result is the dark noise of this wavelength.
None
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength
500, press the button Zero;
2.
Press the button Measure, the result is the noise of 500nm.
Low pressure quartz mercury lamp
1.
Open the lamp cover, put the low pressure quartz mercury lamp into
the lamp seat, and turn it on.
2.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength
546.1;
3.
Press the button Measure, the result is the bandwidth.
22

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Uv-21Uv-31

Table des Matières