8
§ Close the sample chamber by pressing the k key.
§ Press the T key; the following appears on the
display:
"Heater adjustment".
§ Start "Heater adjustment" by pressing the O soft
key.
§ Select "2-point temperature adjustment"; start by
pressing the O soft key.
§ If required, enter the serial number (ref. ID numbe)r
(4). Confirm by pressing the soft key
[Ref-ID]. The reference ID number is a serial
number; for example: 0011901617 (on the
bottom side of the temperature adjustment disk).
§ Press the [Start] soft key to activate the
temperature adjustment routine.
> The heating symbol appears in the display, the
heating time and the previous drying time are
displayed continuously.
§ After you have finished the measurement for
calibrating and adjusting the temperature, you will
hear 3 beeps. Display: Analysis end
§ Press [Info] to display the measured values.
§ Go back by pressing, soft key
> Return to the main menu without performing an
adjustment by pressing the [End] soft key. A
record is printed out automatically.
§ Activate the calibration/adjustment and
temperature adjustment routine on the unit by
pressing the [Adjust.] soft key. This concludes the
adjustment routine. If the final temperature shows
163°C, for example, the temperature is 1°C
outside the tolerance limit of 160°C ± 2°C. This