Factory Calibration Certificate
Example
Test object:
Test equipment number:
Record number: 26/00
Test method:
Measuring points:
Reference
TMD
[°C]
80.69
140.47
202.03
Calibration devices used:
Traceable to: - (Physikalisch-Technische Bundesanstalt (PTB)
German Federal Institute of Physics and Metrology
- German Calibration Service (DKD),
calibration laboratory for temperature measuring equipment
Calibration Certificate for the reference TMD:
- Temperature calibration for the measuring points: 80°C, 140°C, 200°C, 250°C
- PTB emission rate calibration at the temperatures 80°C, 140°C and 200°C
within the wavelength range of 2.5...20 µm
- PTB reflectance calibration at room temperature within the wavelength range
of 0.7...10 µm
Recommended calibration interval: 1 year
Inspector: ______________
Diedrich
Temperature Adjustment Set (YTM03MA)
0011901617
Comparative temperature measurement
80°C, 140°C, 200°C
TMD
Display
Correction
[°C]
[°C]
80.44
0.25
140.58
0.11
201.39
0.64
Reference temperature measuring disk (TMD),
test equipment number: TEM0100
Keithley measuring instrument,
test equipment number: DVM0148
Measurement
Uncertainty
[°C]
0.50
0.50
0.50
Next calibration:
Date: July 19, 2000
Room
Temperature
[°C]
23.3
24.6
25.8
07/01
13