Install CONCAVE feed-in socket
General notes on installation
The feed-in socket may only be installed by a skilled qualified
electrician in accordance with these installation instructions.
Note the details in the manufacturing drawing!
Rear left-hand feed is shown here (rear right-hand feed is mirror
image).
Installation procedure
• Mark out mounting centre line
• Mark out and make wall cut-out relative to edge of medical sup-
ply unit - minimum wall clearance is 10 mm!
• Fit feed-in socket – order no. 20 961 303
• Insert leads for sockets, nurse call, phone etc. as shown in the
manufacturing drawing, allow access length of at least 600 mm
• Install medical gases as shown in drawing; follow correct se-
quence!
Screw-clamped joint
Order no. 20 961 340 - 8 mm
Order no. 20 961 356 - 12 mm
2/5
Centre line (MA)
Lower edge
medical supply
unit
CONCAVE 200
Lower edge
medical supply unit
WA = wall cut-out
CONCAVE 400 double section
Lower edge
medical supply unit
G6
G5
G4
G3
G2
G1
Mat.-Nr. 04882128/4/P- © ILT GmbH 11.06