mmHg LED is fully illuminated and the 200 mmHg LED is at half brightness, the suction level is 175 mmHg. When the 200 mmHg LED
illuminates at full brightness, the unit has reached 200 mmHg.
3. Connect suction tip or catheter as appropriate.
NOTE– If the unit does not maintain vacuum, refer to Troubleshooting.
NOTE– Suction ceases if liquid level reaches float shut-off valve located on underside of canister lid.
CAUTION–Further suctioning may cause damage to the vacuum pump and voids warranty. Equipment service is required if fluid content is
aspirated back into the unit.
Battery Charging
The device is equipped with a factory-installed high capacity rechargeable battery. Connect battery and charge before first use (refer to
Battery Replacement).
1. Connect the unit to either an AC or DC power source.
2. The green external power light should illuminate; the yellow charge indicator will remain lit while the battery is charging.
3. Ensure that the yellow charging light is illuminated when charging begins. As the battery nears a full charge, the yellow battery charging
light may flash for several minutes. This is normal. If your unit does not hold a charge, check that the yellow light turns on when external
power is applied with the power button "Off." If problems persist, contact your medical equipment provider.
NOTE–A fully discharged battery will require up to 5 hours (depending on depth of discharge) of charging to reach full capacity. If unit is not in
use for extended periods, the battery should be recharged a minimum of every 6 months. A fully charged battery will provide approximately
45-60 minutes of continuous operation at zero vacuum level (free flow).
CAUTION–Completely discharging the battery will shorten the battery life. Do not operate the unit for more than a few minutes if the low
battery indicator light is lit. Recharge the battery as soon as possible.
Batter y replacement
1. Using a coin or straight-blade screwdriver rotate latch to unlocked position.
2. Remove door by pulling up on latch.
3. Remove battery from compartment and unplug connector from circuit board.
4. Install new battery by reversing the above steps.
5. Fully charge battery before using.

8. CLEANING INSTRUCTIONS

Collection Canister (not applicable for Jet Compact):
NOTE–Elbow is permanently attached to the lid and cannot be removed.
1. Push power button to turn unit off. Wait for vacuum level to drop.
2. Disconnect external power source from input receptacle on unit (if applicable).
3. Remove canister from unit by disconnecting the patient tubing and filter.
NOTE–The reusable collection canister and lid are single-patient use and should be replaced every 1 to 2 months,depending on use and
cleaning method.
4. Collection canisters and lids should be emptied and thoroughly cleaned after each use. Wash in a hot water/dishwashing detergent
solution and rinse with clean, hot tap water. Then rinse with a commercial disinfectant or one part vinegar to three parts hot water solution.
Thoroughly rinse with hot tap water and air dry. Note: If using a commercial disinfectant, follow manufacturer's recommended instructions
and dilution ratios carefully.
Suction Unit:
1. With the power "Off," disconnect the unit from all external power sources.
2. Wipe the outside housing with a clean cloth dampened with commercial disinfectant/detergent (bacterial-germicidal).
CAUTION– Do not submerge suction unit in water; this will result in damage to vacuum pump.
Tubing:
1. Disconnect tubing from unit.
2. Tubing should be rinsed thoroughly after each use by running tap water through it, followed by a solution of one part vinegar to three
parts hot water.
3. Rinse again with hot tap water and air dry.
4. Keep the outside surface of the tubing clean by wiping with a clean damp cloth.