4) Prior to delivering the orthosis:
Secure the joint bolt (4) and the joint nut (8) with 636K13 Loctite 241.
Verify the proper functionality of the product.
Removing the system bar
1) Unscrew the set screw (9).
2) Tap out and remove the system bar (18).
6 Cleaning
After contact with water containing salt, chlorine or soap, or when they get
dirty, the system joints must be cleaned promptly:
1) Rinse the system joints with clean fresh water.
2) Dry the system joints with a cloth or allow to air dry. Avoid exposure to
direct heat (e.g. from an oven or radiator).
The system joints must be cleaned promptly after coming into contact with
lubricants (e.g. grease, oil):
1) Damp a cloth with a degreasing cleaning agent (e.g. acetone or isopro­
pyl alcohol).
2) Wipe down the system joints or components with the cloth.
7 Maintenance
The manufacturer recommends a semi-annual inspection to verify functional­
ity and check for wear of the system joints.
Spare parts are listed in the section "Components/design".
8 Disposal
Dispose of the product in accordance with national regulations.
9 Legal information
All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
9.1 Liability
The manufacturer will only assume liability if the product is used in accord­
ance with the descriptions and instructions provided in this document. The
manufacturer will not assume liability for damage caused by disregard of this
document, particularly due to improper use or unauthorised modification of
the product.
15