Accessories; Maintenance - Martin Walter Powersonic S Serie Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4.4.

Accessories

Tank cover (stainless steel)
To prevent evaporation of the tank contents and odours nuisance.
Cleaning basket (stainless steel)
For carrying the items to be cleaned. The basket can be suspended in the cleaning tank and it also
can be placed on the edge of the tank with the handles for draining.
Cover with openings for beakers (different sizes)
easy cleaning of small parts
less needs of cleaning solvent
cleaning in different cleaning solvents
The ultrasonic energy is coupled inside the filled beaker by the liquid around through the glass. Fill
the tank with water and add a surface-active agent (e.g. washing-up liquid).
5.

MAINTENANCE

WARNING:
The ultrasonic cleaning unit needs low maintenance.
Regard following notes:
Prior to any emptying or cleaning the unit must be disconnected from mains.
Keep the device clean. Wipe the housing frequently with a damp cloth. Remove cleaning agent residues
immediately.
Ensure that the cleaning liquid does not soil the housing during cleaning.
Clean, replace and dispose the cleaning liquid in accordance with the supplier's recommendations.
Avoid scratching the inner tank surface; this accelerates cavitation (erosion). Ensure the items to be cleaned have
no direct contact with the bottom or inner sidewalls of the tank and try to keep none or only small amounts of
pollution on the bottom of the cleaning tank. If necessary, the cleaning liquid should be drained off prematurely, so
the tank can be cleaned.
Check the unit and the power cord for damages frequently and prior to any use.
A visibly damaged unit or cable must not be put in operation and must be disconnected from the mains.
Do not carry out repairs yourself. Send the unit to the manufacturer for servicing or repair.
There are any parts to maintain or which have to be changed by the user inside the unit.
If the power cord is damaged, it can be removed by unlocking the cap nut (circular plug at rear side of unit). The
power cord must be replaced with the manufacturer's original part only.
0160.0131 / 141219 Crest P 0117_001
Operation
E
Page 25 / 47

Publicité

Table des Matières
loading

Table des Matières