STERILIZATION
WARNING
• Sterilize with autoclave after each patient use.
• Sterilization is required before first use.
• Wear appropriate personal protective equipment during sterilization and lubrication process. This should include eye
protection and gloves.
• Do not use ultrasonic bath or cold sterile/disinfectant chemicals to clean handpiece.
• Do not autoclave handpieces together with other instruments.
• Do not use dry heat sterilization.
• Keep handpiece in sterilization pouch until ready to use.
• Do not cool handpiece manually as damage may occur.
• Do not autoclave at temperatures that exceed 275º F (135º C).
1. Remove bur, then wipe external surface with dry paper towel to remove any excess lubricant and debris. Place paper towel in