TROUBLESHOOTING (continued)
VISUAL
CONDITION OR
DISPLAY
PROBLEM
SEL run
HOLD time-out
E1-E9
Electronics error
LO bAt
Approximately 30 minutes
of battery operation remain
H5
HOLD countdown
LOC
Lock-out disables SET RATE,
SET DOSE and CLEAR VOL
dial settings
8888
Display test
EXPLANATION OR
CORRECTIVE ACTION
Turn dial to RUN.
Remove pump from patient
immediately and contact Abbott
Nutrition.
• Turn dial to HOLD, then RUN.
• Plug power cord into AC outlet
for 12 hours, then continue
feeding.
NOTE: After 30 minutes,
pump alarms continuously
for 1 minute and then shuts
down.
• Turn dial to HOLD.
• If desired, program the HOLD
timer for 1–90 minutes by
depressing the up or down
arrow.
• Turn dial to SET RATE.
• Depress up and down arrows
simultaneously to deactivate.
• Repeat steps to activate.
All segments should be
illuminated.