Troubleshooting - Abbott PATROL ENTERAL PUMP Manuel

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TROUBLESHOOTING

®
The Patrol
Enteral Pump is designed to be highly reliable. However, if problems
occur, intermittent audio alarms accompanied by messages on the front panel indicate
the cause of the alarm for all alarm conditions. The following tips will help correct
problems that may arise:
VISUAL
CONDITION OR
DISPLAY
PROBLEM
no FLO
Flow has stopped due to
occlusion or empty container
doSE FEd
Dose complete—VOL FED is
equal to the SET DOSE
FrEE FLO
Free-flow alarm
SEL rAtE
Dial turned to RUN without
selecting a rate
EXPLANATION OR
CORRECTIVE ACTION
• Turn dial to HOLD.
• Clear obstruction/check
for kinked tubing or empty
container.
• Refill container, if necessary.
• Be sure sight chamber is not
overfilled or covered with a film
of product.
• Turn dial to RUN to start
pump.
• Turn dial to HOLD.
• Turn dial to SET DOSE and
depress up arrow to
increase dose
OR
depress down arrow to zero
the dose setting
OR
turn dial to CLEAR VOL
(after four beeps, volume
fed will be cleared).
• Turn dial to RUN to start
pump.
• Indicates gross uncontrolled
over delivery.
• Remove pump from patient
immediately.
• Verify pump set is properly
installed. Replace set if
necessary.
• Confirm no drops fall when in
HOLD.
• Check accuracy per page 11.
• Turn dial to SET RATE.
• Enter a rate using the up or
down arrow.
13

Publicité

Table des Matières
loading

Produits Connexes pour Abbott PATROL ENTERAL PUMP

Table des Matières