Battery Operation - Abbott PATROL ENTERAL PUMP Manuel

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

BATTERY OPERATION

When pump is unplugged, power is supplied by the internal, rechargeable battery.
1.
If battery if fully discharged and AC power fails during feeding, pump will shut
down and no alarm will sound.
2.
While running on battery power, the display cycles on for 10 seconds and then
off for 10 seconds to conserve power.
3.
Pump can be used while the battery is recharging. Pump automatically charges
the battery when plugged into AC power.
Expected Life and
Charge Time:
1.
When battery is fully charged, pump will operate for 8 hours at 125 mL/hr.
2.
Battery will fully charge in 8 hours with the pump turned off and 12 hours
with the pump turned on.
3.
The battery will charge in any control dial setting including the POWER OFF
position when connected to AC power.
4.
When low battery alarm occurs, the pump will operate for approximately 30
minutes before shutting down.
5.
Battery operating time will diminish with partially charged or older
batteries. Operating on AC power and keeping the pump plugged into an AC
outlet (even while OFF), will help maintain expected battery performance.
Battery Maintenance:
1.
Before initial use on battery power and after extended storage periods, the
pump must be plugged into an AC power source for a minimum of 12 hours.
2.
A battery that has repeatedly been fully discharged and/or left in a
discharged state for an extended period may not recharge properly.
3.
If battery is rarely used, occasionally operating pump on battery power may
extend battery life. The pump should be operated on battery for a minimum
of 6 continuous hours at least once every 6 months for best performance
and battery life.
4.
The pump should be plugged into AC power whenever possible. When not
being used on battery power, pump should ALWAYS remain plugged in to
AC power to maintain battery charge and performance.
Battery Disposal:
Appropriate disposal of batteries is required, and this can be done by an Abbott-
authorized service center. In case of any questions on batteries, please contact the
responsible Abbott representative.