INSTRUCTIONS FOR USE (continued)
TO START PUMP:
12. From the off position, turn pump dial
to SET RATE.
Pump will initiate system self-check
procedure. The audible alarm will
activate twice and 4 "8"s will appear
briefly on the display. If either of these
events does not occur, take the pump
out of service and contact your local
Abbott representative.
13. Select flow rate, from 1 to 300 mL/hr,
by pressing the arrows and observing
the entry on the display. Holding down
an arrow will cause the numbers to
scroll. Never use sharp objects such
as pens, pencils, fingernails or paper
clips to depress the keypad arrows.
14. If desired, turn dial to SET DOSE and
select dose by pressing the arrows.
NOTE: Pump will operate without a
DOSE selected (SET DOSE at 0000).
15. Turn dial to RUN to start feeding.
16. To see feeding volume delivered, turn
dial to VOL FED.
17. To clear the volume delivered, turn
dial to CLEAR VOL. (This occurs
after four beeps.)
18. When feeding is complete, turn
pump dial to OFF.
NOTE: If alarm sounds, turn
pump dial to HOLD. Correct alarm
condition indicated by visual display,
then turn dial to RUN to restart
feeding. (For explanation of alarms,
see pages 13 and 14.)
PRECAUTION: To avoid enteral
product bacterial contamination,
pump set and container should
be replaced as prescribed on
nutritional feeding set labeling. For
single-patient use only. Do not store
for prolonged periods at extreme
temperature.
10