Calibration With Manual Values - XS Instruments 70 Serie Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

The instrument is now ready to recognize the second calibration point.
Next to the string "2ND POINT PH" the different buffers, that the device can recognize automatically,
scroll.
When the 4.01 value is recognized and the icon
string "PRESS OK".
The actual measured value and the Slope% flash on the display; subsequently, the icon of the beaker
pH 4.01
appears next to the green beaker, indicating that the instrument is calibrated in the acid
field.
Remove the electrode, rinse with distilled water and dab gently with paper towel. Dip the sensor in the
pH 10.01 buffer solution ("CHANGE BUFFER").
The instrument is now ready to recognize the third calibration point.
Next to the string "3RD POINT PH", the different buffers, that the device can recognize automatically,
scroll.
When the value 10.01 is recognized and the icon
the string "PRESS OK".
Switching from an acidic to a basic pH may take a few more seconds to achieve stability.
The actual measured value and the second Slope% flash on the display; subsequently, the icon of the
beaker pH 10.01
calibrated in the alkaline field.
At the end of the third calibration point, the instrument returns automatically to measure mode.
To perform a one- or two-point calibration, press the key
Note: electrode calibration is an essential operation for the quality and truthfulness of a
measurement. Therefore, make sure that the buffers used are new, unpolluted and at the same
temperature.
ATTENTION: Before proceeding with the calibration operations, carefully consult the safety data sheets
of the substances involved:
Calibration buffer solutions.
STORAGE solution for pH electrodes.
Filling solution for pH electrodes.
 Calibration with manual values
Example: two-point calibration pH 6.79 and pH 4.65 (DIN 19267)
Access the Setup menu for pH and select in the P1.1  Custom, press twice the key
the measurement and position in pH mode
Press
to enter the calibration mode.
Rinse the electrode with distilled water and gently dab it with paper towel. Dip the electrode in the
first pH buffer solution (e.g. pH 6.79).
Wait for the pH value on the display to stabilize; when the icon
modify it using the directional keys by entering the correct one (e.g. pH 6.79), as suggested by the
string "ADJUST THE VALUE" and by the icon
Note: Check the buffer value according to the temperature.
UM Serie 70Vio EN rev. 1 04.06.2020
appears next to the green and red beakers, indicating that the instrument is
.
.
18
appears, press the key
appears, press the key
once finished the first or second point.
appears and the value flashes,
as indicated by the
, as indicated by
to return to
EN