P3.0
COND SETTINGS
P4.0
TDS SETTING
P8.0
LOG SETTINGS
P9.0
SETTINGS
Temperature measurement ATC – MTC
ATC : The direct measurement of the sample temperature for all parameters is carried out through the
NTC 30KΩ probe, which can be either integrated into the sensor (electrode and / or cell) or external.
MTC : If no temperature probe is connected, the value must be changed manually: keep pressed
or
until the value starts to flash; then adjust it by continuing to use the directional keys;
press
to confirm.

pH Parameter

pH 70 Vio; PC 70 Vio
On this series of devices, it is possible to use pH sensors with integrated temperature probe or to connect
two different sensors. Connect the pH electrode to the BNC type connector marked in green. Connect the
temperature probe to the RCA / CINCH Temp connector always marked with a green background.
The instrument is also able to recognize the DHS sensor, an innovative electrode that stores calibration data
and that can be used immediately after on any enabled instrument.
 pH parameter Setup
In measure mode press
Press the button
UM Serie 70Vio EN rev. 1 04.06.2020
P3.1
P3.2
P3.3
P3.4
P3.6
P3.7
P3.8
P3.9
P4.1
P8.1
P8.2
P9.1
P9.2
P9.3
P9.4
P9.5
P9.6
P9.8
P9.9
to access the SETUP menu.
to access the pH SETTINGS P1.0 menu.
15
Cell Constant
Buffer Selection
Reference Temp
Temp. Compensation Factor
View Cond Cal
Due Cond Cal
Reset Cond Setting
Temp Cond pH
TDS Factor
Data Logger Type
Clear Data
Temperature U.M.
Time and data Set
Backlight mode
Brightness
Sleep Mode
Parameters Setup
Reset