Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
DO NOT block the venting grids of the light source or impede exhaust air
flow. Users must ensure that ventilation air temperatures are below the
maximum rating for the light source and are not influenced by warm
exhaust from other sources.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light
source.
Use only cords provided with the light source.
CAUTION
Plug the AC power cord into a wall outlet using the three-prong plug supplied
with the unit.
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power with
three-prong plug (USA use UL2601-1 rated isolation transformers and/or
power strips only).
When used in clinical or residential areas near radio or TV units,
this equipment may be subjected to radio interference. To avoid
adverse electromagnetic effects, DO NOT operate this equipment near
CAUTION
RF energy equipment.
5.2 CONNECTING THE FIBER OPTIC CABLE
Turn intensity control to the minimum setting. Connect the light cable to the instrument
and then plug the light cable endfitting into the port on the front panel.
LIT158 SUNOPTICS SURGICAL
(English)
®
Rev. B
Date of Revision:07/15/14
Page 9 of 65

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ssl-050

Table des Matières