General technical specifications

90W power supply unit
30W power supply unit
Operating temperature
Operating humidity
Operating altitude
Storage and transport temperature
Storage and transport humidity
Electromagnetic compatibility
Aircraft use
IEC 60601-1 classification
S9 technical specifications
Operating pressure range
Maximum single fault steady state
pressure
Pressure measurement tolerance
Flow measurement tolerance
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 70W (80VA)
Maximum power consumption 110W (120VA)
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 20W (30VA)
Maximum power consumption 36W (75VA)
41ºF to 95ºF (+5ºC to +35ºC)
10–95% non-condensing
Sea level to 8,500' (2,591 m)
-4ºF to 140ºF (-20ºC to +60ºC)
10–95% non-condensing
Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to
IEC60601-1-2, for residential, commercial, and light
industry environments. Information regarding the
electromagnetic emissions and immunity of these
ResMed devices can be found on www.resmed.com, on
the Products page under Service and Support. Click on the
PDF file for your language.
ResMed confirms that the S9 Series meets the Federal
Aviation Administration (FAA) requirements
(RTCA/DO-160F) for all phases of air travel.
Class II (double insulation), Type BF
4 to 20 cm H
O
2
30 cm H
O
2
±0.5 cm H
O ± 4% of the measured reading
2
±6 L/min or 10% of reading, whichever is greater