In a properly functioning, calibrated autoclave steam sterilizer,
effective sterilization can be achieved with the parameters in
Table 1:
Location
Inside the United
States
Outside the
United States
The DirectLink ICP Extension Cable has been tested to withstand
100 autoclave cycles. Cable performance beyond 100 autoclave
cycles has not been evaluated.
Storage
For long term storage of the extension cable, ensure that it is
protected from dust and kept in a clean, dry environment with
these conditions:
Humidity (RH) range: 15% to 85% (non-condensing)
Temperature range: 0° C to +50° C

Service and Repair

The DirectLink ICP Extension Cable is not a repairable product;
please contact your local Integra representative to order a
replacement.

Product End of Life

The cable contains electrical components. Dispose of the
equipment in accordance with local ordinances.
PRODUCT INFORMATION DISCLOSURE
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA") HAS
EXERCISED REASONABLE CARE IN THE SELECTION OF
MATERIALS AND THE MANUFACTURE OF THESE PRODUCTS.
INTEGRA WARRANTS THAT THESE PRODUCTS SHALL
CONFORM TO THE PRODUCT LIMITED WARRANTY AS
PROVIDED IN THE PRODUCT LABELING OR APPLICABLE
PRODUCT CATALOG. THIS WARRANTY IS EXCLUSIVE, AND
INTEGRA DISCLAIMS ALL OTHER WARRANTIES, WHETHER
EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE. INTEGRA SHALL NOT BE LIABLE
FOR ANY INCIDENTAL OR CONSEQUENTIAL LOSS, DAMAGE,
OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING FROM USE
OF THESE PRODUCTS. INTEGRA NEITHER ASSUMES NOR
AUTHORIZES ANY PERSON TO ASSUME ANY OTHER OR
ADDITIONAL LIABILITY OR RESPONSIBILITY IN CONNECTION
WITH THESE PRODUCTS.
Codman, DirectLink and Codman Microsensor are registered trademarks of
Integra LifeSciences Corporation or its subsidiaries in the United States and/
or other countries.
Table 1
Temperature
270° F (132° C) 4 minutes
134° C
5
Duration
Drying Time
30 minutes
3 minutes
30 minutes