4.2

SIMULATION / THERMOCOUPLE OR MILLIVOLT INPUT:

1.
Press key 2 to power up the calibrator.
2.
Press key 5 to be in output mode.
3.
Press key 3 to select the desired type of thermocouple.
4.
Press keys 6, 7, 8, and 9 to define the desired value.
5.
Connect the thermocouple or the voltmeter to output terminal 1
6.
If you want to change the output value, press keys 6, 7, 8, and 9, or change
the type of thermocouple using key 3.
4.3

CONNECTING LEAD :

Use the thermocouple adapter provided as an accessory to make the various
connections desired.
C.A 1621 Thermocouple temperature calibrator
Thermocoulpe Type
K J T B E N R S
-
+
TERMINAL
30V MAX
21