NOTE: When Legal for Trade is set to ON and the
security switch is set to ON, the following menu
settings cannot be changed: Zero Calibration,
Span Calibration, Linearity Calibration, GEO,
Range, Capacity, Graduation, Power On Unit,
Zero Range, Auto Tare, Retain Weight, Legal for
Trade, Stable Range, Auto Zero Tracking, Gross
Indicator, Modes, Units, Stable Only.
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. (See Figures 4 and 5)
Figure 4
Figure 5
CAUTION! Disconnect the unit from the power
supply before cleaning.
PAGE 9 OF 32
OPERATION CONTINUED
Wire Seal
Paper Seal
LOAD CELL SEALING COLLAR
When the scale base is attached to the indicator using
a connector, it is necessary to seal the load cell cable
to the indicator in some jurisdictions. The load cell
sealing collar is available as an accessory.
(See Figure 6)
Figure 6
MAINTENANCE
CLEANING
• Use approved cleaning solutions for the stainless-
steel indicator housing and rinse with water. Dry
thoroughly.
• Do not use solvents, chemicals, alcohol, ammonia
or abrasives to clean the control panel.
Collar Seal
Collar Seal