5 - DC AND AC CURRENTS

Always switch off the circuit to test before connecting the
multimeter to the circuit. If the "Fus." light comes on, change the
faulty fuse(s) (Reminder: minimum voltage of 80 V).
Connect the leads to the multimeter and connect in series to
the circuit with:
- the red lead in the "mA" terminal, up to 500 mA
- the red lead in the "5 A" terminal, from 500 mA to 5 A
Switch off the power supply to the circuit before changing
range.
I When the order of magnitude is not known, place the switch on the
highest range then progressively lower to the appropriate range.
I To get the current in mA or A, multiply the value read on the
appropriate scale by the reading coefficient shown in the table.
5 0 µ A (1)
A D C
S ca le
R e a d in g
(2)
x 1
co e ffic ie n t
V o lta g e d ro p
a t t h e
1 0 0 m V
7 0 0 m V
(3)
te rm in a ls
(4)
A cc u ra cy
P ro te c tio n
Limitation 10 min On, 5 min Off up to +40°C max
.
(1) Common to the 100 mV DC range
(3) Without the leads. Resistance of the pair of leads supplied: approx. 70 mΩ
(4) In % of the end of scale
Do not use the A AC input on unprotected current
transformers.
A A C (1)
5 m A
S cale
R eading
x0,1
coefficien t
Voltage d ro p at
700 m V
the term inals (2)
A ccuracy (3)
Protectio n
Limitation 10 min On, 5 min Off up to +40°C max
(1) Bandwidth: 40 Hz to 5 kHz
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 mΩ
14
5 m A
5 0 m A
5 0 0 m A
5 0 b la ck
x0 .1
x1
x1 0
9 0 0 m V
2 V
2 .5 %
F u s e 0 .5 A H B C
(2) Direct reading in µA
50 m A
500 m A
50 red
x1
x10
800 m V
2 V
2.5 %
Fuse 0.5 A H B C
(3) In % of the end of scale
5 A
x 0 .1
1 .3 V
5 %
F u s e 5 A
H B C
5 A
x0.1
1.3 V
5 %
Fuse 5 A
H BC