Sterilization
Unless otherwise specified, non-sterile products can be re-sterilized using validated steam sterilization methods (ISO 17665 or national
standards).
Ultradent recommendations for packed handpieces and their accessories are as follows:
Cycle type
Sterilization exposure time
Sterilization exposure temperature
Drying time
Saturated steam - forced
4 minutes
132°C
Minimum 20 minutes
air removal (pre-vacuum)
Dry times generally range from 20 to 60 minutes due to the difference in packaging materials, steam quality, device materials, total
mass, sterilizer performance, and varying cool-down time.
The distributor and manufacturer accept no responsibility for sterilization procedures performed by the customer that are not performed
according to these Ultradent recommendations.
Storage
Storage conditions for products labelled "STERILE" are printed on the packaging label. Packaged products should be stored in a dry, clean
environment, protected from direct sunlight, pests, and extremes of temperature and humidity. Use products in the order in which they
are received ("first in, first out" principle), taking into account the expiry date indicated on the label.
J12949 • V1 • (13) • 03/2014 • RB62US020A
Ultradent Products, Inc. • 505 West 10200 South • South Jordan • UT 84095 • USA
www.ultradent.com • customer.support@ultradent.com