Ultradent Ultrawave Manuel D'utilisation page 33

Pièce à main
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Step
Rinsing
Drying
Thermal disinfection
For automated cleaning: thermal disinfection at 90°C for a minimum of 5 minutes.
Inspection
− Handpieces and Nosepieces must be inspected to check that no contamination remains, that they are not corroded, dulled, discoloured
or damaged.
− Before packaging and sterilizing the cleaned products, check they are clean, undamaged and function properly. Repeat cleaning steps
until no visible contamination remains on the device.
− Damaged devices must be discarded. Handpieces must not be lubricated.
Packaging
Use suitable packaging for sterilization, such as a Sterile Barrier System in compliance with ISO 11607. Care should be taken to protect
handpieces from contact with other objects that may damage their surface or the Sterile Barrier System.
Duration (minimum)
2 minutes
40 minutes
J12949 • V1 • (13) • 03/2014 • RB62US020A
Cleaning/decontamination instructions
Rinse with warm distilled or purified water (>40°C)
90°C
11