Transport

6
Transport
Transport in the original packaging or in optionally available
transport box.
7

Storage

Storing the whole system:
Remove respiratory filter and rechargeable battery.
Store the components in a container or cabinet dry and
clean and protect them from direct sunlight and thermal
radiation.
Storing rechargeable batteries:
Deeply discharged batteries may get damaged after prolonged
storage. Charge them to 50 to 70 % prior to storage.
If storage lasts for over 6 months, charge them in the
meantime.
Do not store rechargeable batteries for prolonged periods
outside the recommended temperature range. This might
reduce the remaining capacity and number of potential
charge cycles.
8

Disposal

Dispose of the device according to local waste disposal
regulations.
9

Technical data

9.1

Overall system

Flow rate of respiratory protective
device/helmet/visor
Flow rate of full facepiece
Rated period of service
1
Operating temperature
Working and storage humidity
1)
Storage temperature
Noise
International Protection Code
RFID
Technology
Frequency range
Broadcast transmission capacity
14
170/190/210 l/min
115/130/145 l/min
4 hours
with standard
rechargeable battery
8 hours
with long-life
rechargeable battery
14 °F to 140 °F (-10 °C
to 60 °C)
1)
≤ 95 % relative humidity
-4 °F to 140 °F (-20 °C to
60 °C)
approx. 64 dB(A)
IP 65
Induction transfer
13553 to 13567 kHz
-2,30 dBµA/m (10 m)
Bluetooth
Technology
Frequency range
Broadcast transmission capacity
1
For the battery charger and rechargeable batteries, refer to the
separate information provided in this chapter.
For other components, refer to the corresponding Instructions for
Use.
9.2

Rechargeable batteries

Operating temperature
Operating/storage area humidity
Storage temperature
Charging temperature
Standard rechargeable battery
Charging time
Operational life time after a full
charge
Rated voltage
Rated capacity
Stored energy
Long-life rechargeable battery
Charging time
Operational life time after a full
charge
Rated voltage
Rated capacity
Stored energy
1
Varies depending on the preset flow rate and the employed respi-
ratory filter and respiratory inlet covering type
9.3

Standard battery charger

Input voltage
Input current
Output voltage
Output current
International Protection Code
Operating temperature
FHSS 2.4 GHz
(BT 2.1 + EDR)
2402.0 to 2483.5 MHz
0,97 mW / -0.14 dBm
EIRP
14 °F to 131 °F (-10 °C to
55 °C)
≤ 95 % relative humidity
-4 °F to 122 °F (-20 °C to
50 °C)
32 °F to 122 °F (0 °C to
50 °C)
< 4 hours
1)
approx. 4 hours
10,8 V
3,35 Ah
36 Wh
< 4 hours
1
approx. 8 hours
10,8 V
6,70 Ah
72 Wh
15 V
4 A
9 - 12,6 V
4 A
IP 30
32 °F to 122 °F (0 °C to
50 °C)
X-plore 8000