EMC data
Guidance and manufacturer's declaration - electromagnetic immunity - for equip-
ment and system that are not life-supporting. The model BW-CX10 is intended for
use in the electromagnetic environment specified below. The customer or the user of
the model BW-CX10 should assure that it is used in such an environment.
Immunity
EN 60601
test
test level
3 Vrms
Conducted
150 kHz
RF EN
to 80
61000-4-6
MHz
3 V/m
Radiated
80 MHz
RF EN
to 2.5
61000-4-3
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy. To assess the electromag-
netic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the InfraredThermom-
eter type BW-CX10 is used exceeds the applicable RF compliance level above, the Infrared
Thermometer type BW-CX10 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or
relocating the Infrared Thermometer type BW-CX10.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
44
Compliance
level
Portable and mobile RF communications equipment should be
used no closer to any part of the Infared Thermometer type BW-
N/A
CX10, including cables, than the recommended separation dis-
tance calculated from the equation applicable to the frequency of
the transmitter. Recommended separation distance
d = [ 3.5
V1
d = [ 3.5
E1
d = [ 7
E1
Where P is the maximum output power rating of the transmitter in
3 V/m
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance
level in each frequency range. b Interference may occur in the vi-
cinity of equipment marked with the following symbol:
BewellConnect - BW-CX10 - User Manual - 042017v3
Electromagnetic
environment - guidance
] √
P
150 MHz to 80 MHz
] √
P
80 MHz to 800 MHz
800 MHz to 2.5 GHz
] √
P