Invacare Perfecto 2 Serie Manuel D'utilisation page 15

Concentrateurs à oxygène avec senso2 compatible homefill
Masquer les pouces Voir aussi pour Perfecto 2 Serie:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 33
Weight:
Shipping Weight:
Operating
Ambient
Conditions:
Exhaust Cooling
Air Temperature:
Oxygen Output
Temperature:
Cabinet:
No AP/APG
Standards and
Regulatory Listing:
CE marked model
Electrical:
Placement:
Tubing:
Time of
Operation:
Recommended
Storage and
Shipping
Temperature:
Environment:
Minimum
Operating Time:
Part No 1163145
SECTION 5—TYPICAL PRODUCT PARAMETERS
Perfecto
AW - 20.5 kg ± 1 kg (45 lbs ± 2 lbs)
2
Perfecto
VAW - 18.1 kg ± 1 kg (40 lbs ± 2 lbs)
2
Perfecto
AW - 22.7 kg ± 1 kg (50 lbs ± 2 lbs)
2
Perfecto
VAW - 20.4 kg ± 1 kg (45 lbs ± 2 lbs)
2
10°C - 35°C (50°F - 95°F) at 20-60% relative humidity
Less than Ambient +19°C (+ 45°F)
Less than Ambient +3°C (+ 8°F)
Impact Resistant flame-retardant plastic cabinet that conforms to
UL 94-V0.
Not suitable for use in the presence of a flammable anaesthetic
mixture.
IEC/EN61000-3-2
IEC/EN61000-3-3,
IEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2
ISO8359
MDD 93/42/EEC, Annex I and IX
IRC5PO2AW / IRC5PO2VAW
No extension cords.
No closer than 30.5 cm (12 in) from any wall, furniture,
draperies, or furniture to assure sufficient air flow. Avoid deep
pile carpets and heaters, radiators or hot air registers. Floor
location only.
2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-Proof
Tubing (DO NOT pinch).
Up to 24 hours per day.
-29°C to 65°C (-20°F to 150°F) at 15-95% relative humidity.
Smoke and soot-free. No confined spaces (Example: No closets).
30 Minutes
15
Perfecto
™Series
2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Perfecto 2 irc5po2awPerfecto 2 irc5po2vaw

Table des Matières