Do not use fluids exceeding 80 °C.
There is no specific advice on when to replace the cleaning solutions as solutions life will
depend on the volume of items being cleaned and how contaminated they are to start with.
We recommend removal of the solution from the bath when it is not to be used for storage
in a separate container, filtering at this stage may be beneficial to prolong solution life. Fluid
removal allows for the routine cleaning of the bath where an accumulation of contaminants
such as grease, dirt, swarf and oil can be removed from the base of the bath. If left the build-
up of hard contaminants can rub on the base of the bath and along with the ultrasonic action
this can damage the surface of the base of the tank.
If any of the electrical connections become wet or damp the Cleaner should be isolated at the
supplies and checked and rectified by a competent person before reuse.
Cleaning cycle times will vary dependant on the item being cleaned, the solution being used
and the finish required, typically we would recommend a 5-minute cycle as a starting point
and increase the cycle time accordingly until the desired results are achieved.
The maximum cycle run time must be limited to 45 minutes and should be followed with a 20
minutes rest period to ensure long and reliable operation.
Never drop an item onto the base of the bath.
A Degass cycle is recommended at the beginning of each day or when the solution has been
changed, the purpose of the Degass cycle is to purge excess pockets of air from the solution
which will ensure the cleaning results are more uniform and reliable. A Degass cycle can be
performed by operating the Pulse, or full power cleaning cycle for approximately 10 minutes.
If you forget to Degass it is likely that the cleaning action is likely to be significantly reduced
on your first clean.
If at any point a defect is noticed with the Ultrasonic cleaner, leads or generator/s isolate the
Cleaner from the supply and await advise from a competent appointed person, spare parts
are available by contacting us directly. Engineers are also available to offer help and support
you with technical queries and advice.
Care should be taken when using the heaters, check the thermostat position before
powering on the cleaner. Do not leave unattended, the heating circuit is not intended for
continuous use, it should be powered approximately 1 hour before the cleaner is to be used
and switched off at the end of each working day. Failure to do this will reduce the elements
service life and may lead to premature failure, not covered under warranty.
Only responsible adults should operate ultrasonic cleaners.
Do not place hands or fingers in the bath. Contact exposure to ultrasonic cavitation is
suspected to cause living tissue and bone damage.
4
ULTRASONIC CLEANER USER MANUAL
www.rswww.com/- Images & Content @2013 RS Components Ltd