Maintenance
Your mini-centrifuge is maintenance-free. Clean unit only when it is not plugged into an
electrical outlet. When necessary, the housing and rotor can be wiped using a damp cloth and
a mild, non-corrosive detergent.
WARNING
Do not use any solvent on the unit that may attack plastic or cause cracks in the
rotor. Ensure the rotor is protected from corrosion and mechanical damage. The rotor
must be cleaned with a neutral cleaning liquid.
If hazardous sample is spilled on the unit and or leaked into the unit, do not run the
centrifuge unless it is safe to do so.
Dissassembly of Rotors for Cleaning
1.
Remove rotor from unit
2.
Remove screws from the bottom of the rotor
3.
Clean as needed
4.
Reassemble rotor, making sure that the alignment tab is engaged into the slot.
5.
Test run reassembled empty rotor to ensure it is still balanced
NOTICE
Ensure all parts are thoroughly dried prior to operation.
Hazardous Chemicals:
1.
Cleaning and decontamination may be necessary as a safeguard before laboratory
centrifuges, rotors, and any accessories are maintained, repaired, or transferred. The user
must follow any and all instructions provided in the manual pertaining to this process.
2.
In the event hazardous material is spilt, the user is responsible for carrying out the
appropriate decontamination procedure.
3.
Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed methods will not
damage the equipment.
Ordering Information
Mini-centrifuge with AC adapter, 2 rotors, rotor cover, and tube adapters
Standard Rotor
Strip Tube Rotor
Rotor Knob
Standard Rotor Cover
Cord Set
Power Adapter
Tube Adapters (0.2 mL x6, 0.5 mL x6)
12
HS10050
HS100503
HS100504
HS100505
HS100506
HEA100502
HEA120527
x2 HEA120015