Surgical Clipper Specifications:
Mode of Operation
BD Disposable Blades
Mode of Operation
Operating Voltage / Current
Weight
Dimension
Full Charge Time
Continuous Run Time
(Fully Charged)
Operating Noise
Type of Protection Against
Electric Shock
Degree of Protection Against
Electric Shock
Degree of Protection Against
Flammable Anesthetic
Ambient Operating Temperature,
Humidity, and Atmospheric Pressure
Storage Temperature, Humidity,
and Atmospheric Pressure
Continuous
Single Use
3�25V DC / 390mA (max�)
121 grams (4�3 oz�)
6�0" (153 mm) H × 1�9" (49 mm) W ×
1�5" (39 mm) D
Maximum 17 hours
Disposable Blades REF 4406 & REF 4403A:
Minimum of 80 minutes continuous run-
time up to 120 minutes.
Disposable Blade REF 4412A: Minimum of
50 minutes continuous run-time.
REF 4406 Blade is less than 72dB�
REF 4412A Blade is less than 69dB�
REF 4403A Blade is less than 75dB�
Internally Powered
Type B Applied Part
Equipment not suitable for use in the
presence of a flammable anesthetic�
+41 °F (+5 °C) to +95 °F (+35 °C), 30%
to 75% RH, 70 to 106 kPa�
+14 °F (-10 °C) to +140 °F (+60 °C),
10% to 100% RH, 50 to 106 kPa�
Surgical Clipper Specifications (continued):
Waterproofing
Battery Type
Battery Life
Charging Adapter Specifications:
Weight
Power Source
Type of Protection Against
Electric Shock
Dimension
Ambient Operating Temperature,
Humidity, and Atmospheric
Pressure
Storage Temperature, Humidity,
and Atmospheric Pressure
Waterproofing
Cord Plug Type
IPX7 Rating
Li-ion 680mAh
Minimum 350 Cycles
200-250g ± 10g� (7-8�80 Oz�)
100–240 V AC; 50–60 Hz, 30mA
Class II
3�15" (80 mm) H × 2�40" (61 mm) W ×
1�73" (44 mm) D & Cord Length 73�62"
(1870 mm)
+41 °F (+5 °C) to +95 °F (+35 °C), 30%
to 75% RH, 70 to 106 kPa�
+14 °F (-10 °C) to +140 °F (+60 °C),
10% to 100% RH, 50 to 106 kPa�
IPX7 Rating
5514A - Type B
5514E - Type F
5514K - Type I
5514U - Type G