OPTION 011 DATA OUTPUT
INSTALLATION INSTRUCTIONS FOR OPTION 011 DATA OUTPUT FOR AE BALANCES
General information
By using an Option 011 Data Output, a METTLER TOLEDO AE balance can be connected to one or several of the following
instruments.
– METTLER TOLEDO CL instruments, METTLER TOLEDO GA40 Printer (with transfer button at the 011 data output)
– Non-METTLER TOLEDO instruments with active 20 mA data interfaces
– Non-METTLER TOLEDO instruments with RS232C data interfaces (computers, terminals)
The METTLER TOLEDO brochure "The Option 011 Data Output for METTLER TOLEDO AE balances (CL/RS232C)", order No.
701091, provides information about the data traffic with non-METTLER TOLEDO instruments, as well as hardware definitions.
Subject to technical changes and to the availability
of the accessories supplied with the instruments.
Technische Änderungen und Änderungen im
Lieferumfang des Zubehörs vorbehalten.
Sous réserve de modifications techniques
et de disponibilité des accessoires.
Reservadas las modificaciones técnicas
y la disponibilidad de los accesorios.
© Mettler-Toledo GmbH 2000
701202 Printed in Switzerland 0006/33.33
Mettler-Toledo GmbH, Laboratory & Weighing Technologies, CH-8606 Greifensee, Switzerland
Phone +41-1-944 22 11, Fax +41-1-944 30 60, Internet: http://www.mt.com
1
38750
Opening the instrument
This is required only if it is necessary to select a configuration that is
different from the way the instrument has been set at the factory.
(Factory setting corresponds to the standard METTLER TOLEDO CL
configuration: 2400 baud/even parity/send on transfer. Cable shielding
RS232C: not grounded).
– To open the instrument, loosen the four screws in the corners
(arrows).
Configuring the data output
– Baud rate: according to the setting of the instrument(s) to be
connected, set with coding plug (1) according to instructions on P/C
board (factory setting 2400 baud).
– Operating modes: Before being shipped from the factory, all
positions of the operating mode switch (2) are set to OFF.
Position
1)
2)
3)
4)
– Cable shielding (RS232C): Plug in coding plug (3) at center and
right (4) → cable shielding is at housing and ground.
Plug in coding plug (3) at center and left (5) (or use no coding plug
at all) → cable shielding is floating (not grounded).
ON
OFF
parity disable
parity enable
odd parity
even parity
[with 1) OFF]
[with 1) OFF]
space
mark
[with 1) ON]
[with 1) ON]
no function
no function
send continuous
send on transfer
P701202