IDEAL 61-357 Manuel D'instructions page 23

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Function
Fuse Notification
Low Battery
Indication
APO
Restore APO
Disable Auto
Power Off
Function
Mechanical
Housing
Description
The test leads CANNOT be short circuited or used for measuring while
testing the fuse.
1. If the 600mA/1000V fuse is blown, setting the rotary switch to the
"mA" or "uA" setting, the LCD will display "FUSE" and "
the buzzer will beep continuously at the same time.
2. If the 11A/1000V fuse is blown, setting the rotary switch to the "A"
setting, the LCD will display "FUSE" and "
beep continuously at the same time.
When the battery voltage < 3.6 ±0.2V, the low battery indication '
is displayed on the screen and the meter will still work normally. When
the battery voltage drops to less than 3.1 ±0.2V, "bAtt" is displayed for
5 seconds then shuts off. When the battery voltage is less than 3.2V,
accuracy is no longer assured.
The unit will be automatically power off after 30 minutes of inactivity
and enter the low-power state. Current draw is approx. ≤50 micro A.
All the buttons can wake up the unit, or rotate the dial switch to the OFF
setting and then turn the unit back on to wake it up.
Pressing the "Select" key while turning on the unit on at the same time,
will cancel the auto shutdown function. Buzzer will beep 5 times and the
LCD will not display the "APO" symbol.
Single Injection Molding with rubber boot.
", and
.", and the buzzer will
'
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

61-347

Table des Matières