Step
Settings
1.
O/P off
2.
0V
3.
Max V, 100mA
4.
Max V, 100mA
5.
I pot at 0
6.
2V, 0mA
7.
2V, 0mA
8.
2V, max I
9.
2V, max I
10.
2V, 0mA
11.
2V, 0mA
12.
2V, 500mA
13.
2V, 500mA
14.
O/P off
The Manufacturers or their agents overseas will provide a repair service for any unit developing a
fault. Where owners wish to undertake their own maintenance work, this should only be done by
skilled personnel in conjunction with the service manual which may be purchased directly from
the Manufacturers or their agents overseas.
Cleaning
If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent.
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER
ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER
CLEAN WITH SOLVENTS.
40
Display
Adjust for
Cal count
0V ± 2mV
Cal count
0V ± 2mV
Cal count
15/30/60V ± 2mV
O/P volts
Display = DVM
I pot
Display = 0
Cal count
0mA ± 0.2mA
O/P amps
Display = milli-ammeter
Change load
Cal count
1.5/3A ± 1mA
O/P amps
Display = ammeter
Cal count
0mA ± 0.02mA
O/P amps
Display = milli-ammeter
Cal count
500mA ± 0.1mA
O/P amps
Display = ammeter
'End'
Load
DVM
DVM
DVM
DVM
milli-ammeter
milli-ammeter
Ammeter
Ammeter
milli-ammeter
milli-ammeter
ammeter
ammeter
-
Detail
O/P off volts
V DAC zero
V DAC scale
V metering scale
-
I pot ADC zero
High I DAC zero
High I metering zero
High I DAC scale
High I metering scale
Low I DAC zero
Low I metering zero
Low I DAC scale
Low I metering scale
-
Stores and exits
Maintenance