SYMPTOM
Alarm:
Rapid
Beep...Beep...
Beep...Beep
Alarm:
YELLOW or RED light Illuminated.
Alarm: Continuous On RED light
* Only applies to
IRC5PO2
GREEN light with YELLOW Light
Flashing.
1148074-F-06
PROBABLE CAUSE
Potential Obstruction Alert
1.
Possible internal obstruction in the
oxygen path. Kinked or blocked
tubing, cannula or humidifier.
2.
Flowmeter set at 0.5 L/min or less.
System Failure:
1.
Low oxygen purity.*
2.
Kinked or blocked tubing, cannula
or humidifier.*
3.
Flowmeter set at 0.5 L/min or less*.
4.
Unit overheating due to blocked
air intake.
5.
Insufficient power at outlet.
6.
Internal repairs required.
1.
Internal repairs required.
SOLUTION
1.
Inspect for kinks or blockages. Correct, clean or replace
item. Once corrected, turn power Off for 60 seconds and
then turn power back ON.
2.
Reset flowmeter to prescribed flowrate. If prescribed
flowrate is below 0.5 L/min, use the Pediatric Flowmeter
IRCPF16 (Only with Perfecto2 5).
The use of the pediatric flowmeter will deactivate the
Potential Obstruction Alert.
1.
Clean or Replace filters.
2.
Inspect for kinks or blockages. Correct, clean or replace
item. Once corrected, turn power Off for 60 seconds and
then turn power back On.
3.
Check flowmeter is set to 1.0 L/min or more. Refer to
Typical Product Parameters.
4.
Do one or both of the following:
a.
Remove and clean cabinet filters.
b.
Move concentrator at least 12 in (30,5 cm) from walls,
draperies, and furniture.
5.
DO NOT use extension cords. Move to another electrical
outlet or circuit.
6.
Call Service Provider.
1.
Call Service Provider.