Battery Charge Indicator in Charging state:
After 30 minutes of being off the charger the SkinPen
to remind users that it needs to be placed on the charger. This notification may
be cleared by either simply powering on the device, or by placing the device on
the charger.
• When the SkinPen
1, 2, 3 and charger indicator LED will blink sequentially for a
3
2
second to indicate it is connected with the charger base.
1
• When the SkinPen
base the LED's stay solid which indicates the charge level in
the battery. They then go off to maintain battery longevity.
Battery
Battery
Charge ≤ 30%
Charge = 100%
3
• Battery Charge = 100%:
2
LED 1, 2, 3 OFF to maintain battery longevity.
1
3
• Battery Charge > 90%:
2
LED 1, 2, 3 ON.
1
70% <
Battery
Charge
3
• 70% < Battery Charge ≤ 90%:
2
LED 1, 2 ON, LED 3 repeat on/off 1 sec.
30% <
1
Battery
Charge
3
• 30% < Battery Charge ≤ 70%:
2
LED 1 ON, LED 2 repeat on/off 1 sec.
1
3
• Battery Charge ≤ 30%:
2
LED 1 repeat on/off 1 sec.
1
14
PN#16130015 | Rev. C | International
Precision will beep
®
Precision is placed on the charger, LED
®
Precision is removed from the charger
®

13. SPECIFICATIONS

Technical Information of SkinPen Precision
Product Name
SkinPen
Precision Handpiece
®
Model Number
SkinPen
Precision Charger Base
®
Model Number
Bellus Medical FDA Registration #
FCC ID
Weight and Unit
Electrical Requirements
Output voltage:
Charger Time
Working Time
Speed
Needles
Operation
AC Adapter
Charger base transmitting frequency
®
SkinPen Precision
®
100
101
3010392991
2AGLK-101
≤ 5oz /155mm length and
max. outer diameter of 34mm
Charger Base Input: 5VDC, 2A max
5W (max)
From 10% charge to 90%
Charge within 14 hours
> 6 hours
(under normal use conditions)
6300RPM – 7700RPM
• 14 total solid needles
• 32 BWG (gauge)
• <32 RMS (roughness)
• Medical grade Stainless Steel
• EU RoHS compliant
• Sharpness specification within the
Radius 0.005mm (Max)
• Maximum extension of the needles
from the needle head surface is less
than 2.7mm
Cordless
Medical Grade, Universally
compatible power requirements:
100-240 VAC at 50-60 Hz
Between 110kHz and 205kHz
with efficiency around 73%
PN#16130015 | Rev. C | International