Description of your unit's symbols:
There are a number of technical markings on your unit. These
can be explained as follows:
Inside the battery compartment '+' indicates positive polarity
and '-' indicates negative polarity. The batteries' should be
positioned as indicated by three embossed outlines in the
base of the compartment.
The unit and belt are made in China for Bio-Medical Research
Ltd., Parkmore Business Park West, Galway, Ireland.
The unit requires three 1.5 AAA volt DC batteries. DC is
indicated by the symbol:
This symbol means "Attention, consult the
!
accompanying documents".
This symbol means type BF equipment.
This symbol on your unit is to indicate conformity to
the requirements of the Medical Device Directive
0366
(93/ 42/EEC). 0366 is the number of the notified
body (VDE).
SN stands for "serial number". Inside the box is the serial
number specific to this unit. The letter preceding the number
indicates the year of manufacture, where 'S' denotes 2013,
'T' denotes 2014 etc.
The batch number of the belt is represented on the belt pack-
aging, by the number corresponding with the LOT symbol.
www.slendertone.com
English
P
P
R O G R A M
R O G R A M
• Output Current
• Pulse Rate
• Pulse Duration (Width)
• Timer
• Compliance
• Waveform
• Maximum Charge Per Pulse
• Power Source
• Tolerances
P A R A M E T E R S
P A R A M E T E R S
0-70 milliamps, adjustable (up to 1K ohms load)
50, 55, 60, 70, 75 and 99 pulses per second (PPS)
Positive pulse width 200–350 microseconds / Negative pulse width
200–350
• Mode # 1
Pulse rate 70 PPS / On time = 2.5 seconds / Off time = 2.5 seconds
Ramp Up and Down = 2 seconds / Timer = 20 minutes
• Mode # 2
Pulse rate 60 PPS / On time = 3.5 seconds / Off time = 3.5 seconds
Ramp Up and Down = 2 seconds / Timer = 25 minutes
• Mode # 3
Pulse rate 50 PPS / On time = 4.5 seconds / Off time = 4.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
• Mode # 4
Pulse rate 50 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 2 seconds / Timer = 40 minutes
• Mode # 5
Pulse rate 55 PPS / On time = 6.5 seconds / Off time = 6.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
• Mode # 6
Pulse rate 55 PPS / On time = 7.5 seconds / Off time = 7.5 seconds
Ramp Up and Down = 1-2 second / Timer = 30 minutes
• Mode # 7
Pulse rate 55-75 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 0.4-2 seconds / Timer = 35 minutes
• Mode # 8
Pulse rate 55-75 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 0.4-2 seconds / Timer = 40 minutes
• Mode # 9
Pulse rate 70 PPS / On time = 3 seconds / Off time = 3 seconds
Ramp Up and Down = 2 seconds / Timer = N/A
• Mode # 10
Pulse rate 99 PPS / On time = 2 seconds / Off time = 2 seconds
Ramp Up and Down = 2 seconds / Timer = N/A
20, 25, 30, 35 and 40 minutes depending on the mode.
Complies with 60601-1
Symmetrical biphasic with zero net DC
15 microcoulombs under normal conditions
3 x 1.5V AAA Batteries
All electrical specifications are ± 10% into real 500-ohm load
microseconds / Inter-phase interval 100 ± 40 microseconds
15